Table 8.
Blood Glucose Controller Performance Incorporating Stochastic Model Generated from Fitted Insulin Sensitivity Profiles for Patients in Table 4
| No stochastic forecasts | Stochastic model | |
|---|---|---|
| % measurements within 72–126 mg/dl | 76% | 76% |
| % measurements <72 mg/dl | 3.4% | 3.2% |
| % measurements <47 mg/dl | 0.45% | 0.34% |
| Num. patients with a measurement <47 mg/dl | 4 | 4 |
| BG median | 108 | 108 |
| BG IQR | 97–122 | 97–122 |
| LBG-P ratio | 0.59 | 0.44 |