Skip to main content
. 2009 Sep;3(5):1066–1081. doi: 10.1177/193229680900300510

Table 8.

Blood Glucose Controller Performance Incorporating Stochastic Model Generated from Fitted Insulin Sensitivity Profiles for Patients in Table 4

No stochastic forecasts Stochastic model
% measurements within 72–126 mg/dl 76% 76%
% measurements <72 mg/dl 3.4% 3.2%
% measurements <47 mg/dl 0.45% 0.34%
Num. patients with a measurement <47 mg/dl 4 4
BG median 108 108
BG IQR 97–122 97–122
LBG-P ratio 0.59 0.44