Table 3. Figures of merit.

Ionization mode Sensitivity Selectivity Applicability Multimode use
ESI fmol n/aa peptides, proteins yesc
APCI pmol n/ab small molecules yesd
CD-ECI zmol yes ferrocene labeled biomolecules, small molecules yesc
a

CD-ECI achieves selectivity according to the ionization potential of analytes vs. capillary material, ESI and APCI do not have this option

b

under standard HPLC-MS conditions; it should be noted though that APCI can be made selective by changing the nature of the ion-molecule chemistry

c

ESI and CD-ECI are interchangeable without hardware change

d

Hardware change (APCI probe) necessary