Skip to main content
. Author manuscript; available in PMC: 2010 Nov 1.
Published in final edited form as: Anal Chem. 2009 Nov 1;81(21):9086–9095. doi: 10.1021/ac9017692

Table 1.

Channel dimensions for the device shown in Fig. 1A. Fluidic resistances (R) were defined by the channel cross-sectional area, length, shape, and fluid viscosity. The viscosity values used for water and silicone oil were 0.89 × 10−6 and 17.8 × 10−6 kPa s, respectively.

Resistor Length (mm) Width (mm) Depth (mm) Resistance (kPa s mm−3)
25 ° C 37 °C
Rislet* 1.70 / 0.50 0.30 / 0.05 0.125 / 0.035 4.41 3.40
Raq 39.50 0.05 0.035 345 266
Roil 4.65 0.09 0.035 344 265
Rout 22.00 0.09 0.035 1630 1260
*

Values for Rcal are equivalent to values shown for Rislet.