Skip to main content
. 2007 Mar;1(2):205–210. doi: 10.1177/193229680700100211

Table 2.

Probability of Insulin Dose Errora

Meter Blood sample number Coding % probability of occurrence

± 1 unit ± 2 units ± 3 units ± 4 units ± 5 units
A 139 Correctly coded (autocoding) 27.65 0.96
B 140 Correctly coded (autocoding) 35.42 1.41
C 143 Correctly coded (manual coding) 43.07 6.57 0.48
C 143 Miscoded 1 40.20 40.66 14.00 1.4 0.04
C 143 Miscoded 2 36.98 4.22 0.22
D 136 Correctly coded (manual coding) 32.55 1.77 0.04
D 136 Miscoded 1 49.61 28.29 5.92 0.78
D 135 Miscoded 2 41.85 15.64 0.52
E 137 Correctly coded (manual coding) 44.64 7.06 0.49
E 138 Miscoded 1 33.99 50.00 15.14 0.40
E 137 Miscoded 2 26.7 49.7 22.25 1.21
a

Calculated insulin dose error for all meters across all time points based on the low-dose algorithm and on a normal distribution assumption.