Table 2.
Meter | Blood sample number | Coding | % probability of occurrence | ||||
---|---|---|---|---|---|---|---|
± 1 unit | ± 2 units | ± 3 units | ± 4 units | ± 5 units | |||
A | 139 | Correctly coded (autocoding) | 27.65 | 0.96 | |||
B | 140 | Correctly coded (autocoding) | 35.42 | 1.41 | |||
C | 143 | Correctly coded (manual coding) | 43.07 | 6.57 | 0.48 | ||
C | 143 | Miscoded 1 | 40.20 | 40.66 | 14.00 | 1.4 | 0.04 |
C | 143 | Miscoded 2 | 36.98 | 4.22 | 0.22 | ||
D | 136 | Correctly coded (manual coding) | 32.55 | 1.77 | 0.04 | ||
D | 136 | Miscoded 1 | 49.61 | 28.29 | 5.92 | 0.78 | |
D | 135 | Miscoded 2 | 41.85 | 15.64 | 0.52 | ||
E | 137 | Correctly coded (manual coding) | 44.64 | 7.06 | 0.49 | ||
E | 138 | Miscoded 1 | 33.99 | 50.00 | 15.14 | 0.40 | |
E | 137 | Miscoded 2 | 26.7 | 49.7 | 22.25 | 1.21 |
Calculated insulin dose error for all meters across all time points based on the low-dose algorithm and on a normal distribution assumption.