. 2008 Jul-Sep;33(3):100–107. doi: 10.4103/0971-6203.42750

Table 3.

Output ratio (OR) for square fields in full scatter water phantom for Theratron Phoenix Co-60 compared with ESTRO Booklet No:6 published data for MDS Co-60 Theratron 780

Side of square field (cm) Theratron Phoenix Co-60 MDS Co-60 Theratron 780 % Deviation
5 0.857 0.858 −0.12
6 0.887 0.897 −1.11
8 0.949 0.951 −0.21
10 1.000 1.000 0.00
15 1.081 1.093 −1.10
20 1.142 1.152 −0.87
25 1.183 1.200 −1.42
30 1.193 1.225 −2.61
35 1.204 1.236 −2.59