Table 7.
Phantom Scatter correction factor (Sp) of Theratron phoenix Co-60 for square fields compared with S and G data and MDS Co-60 Theratron 780 ESTRO Booklet No:6 published data
| Square field size (X/Y) (cm) | Theratron phoenix | S and G Data Co-60 | MDS Co-60 Theratron 780 | % Deviation with S and G data |
|---|---|---|---|---|
| 5 | 0.884 | 0.871 | 0.898 | 1.49 |
| 6 | 0.912 | 0.904 | 0.918 | 0.88 |
| 8 | 0.962 | 0.945 | 0.961 | 1.80 |
| 10 | 1.000 | 1.000 | 1.000 | 0.00 |
| 15 | 1.053 | 1.068 | 1.069 | −1.40 |
| 20 | 1.083 | 1.104 | 1.110 | −1.90 |
| 25 | 1.109 | 1.135 | 1.150 | −2.29 |
| 30 | 1.118 | 1.150 | 1.166 | −2.78 |
| 35 | 1.134 | 1.165 | 1.186 | −2.66 |
PMMA - Polymethyl methacrylate