Skip to main content
. 2008 Jul-Sep;33(3):100–107. doi: 10.4103/0971-6203.42750

Table 7.

Phantom Scatter correction factor (Sp) of Theratron phoenix Co-60 for square fields compared with S and G data and MDS Co-60 Theratron 780 ESTRO Booklet No:6 published data

Square field size (X/Y) (cm) Theratron phoenix S and G Data Co-60 MDS Co-60 Theratron 780 % Deviation with S and G data
5 0.884 0.871 0.898 1.49
6 0.912 0.904 0.918 0.88
8 0.962 0.945 0.961 1.80
10 1.000 1.000 1.000 0.00
15 1.053 1.068 1.069 −1.40
20 1.083 1.104 1.110 −1.90
25 1.109 1.135 1.150 −2.29
30 1.118 1.150 1.166 −2.78
35 1.134 1.165 1.186 −2.66

PMMA - Polymethyl methacrylate