Energy dependence of Al2O3:C (0.3 mm thick and 7 mm diameter, wrapped in black tape of thickness 34 mg/cm2) to photon (at dmax) and electron radiotherapy beams from Varian 21 EX linear accelerator[65]

Radiotherapy beams Relative response Corrected relative response
6 MV Photon 1.000 1.000
18 MV Photon 1.005 1.005
9 MeV Electron 1.019 1.000
20 MeV Electron 1.023 1.003