Skip to main content
. 2009 Sep 9;47(11):3682–3691. doi: 10.1128/JCM.00781-09

TABLE 3.

Intra-assay and interassay reproducibility analysis of triplex HTLV and singleplex β-globin qPCR assays, based on standard templates

Analysis group, HTLV type, and dilution μ ± σ %CV
Intra-assay reproducibility
    HTLV-1
        105 22.02 ± 0.08 0.38
        104 25.31 ± 0.07 0.26
        103 28.84 ± 0.14 0.47
        102 32.33 ± 0.17 0.53
        10 36.63 ± 0.66 1.81
    HTLV-2
        105 21.94 ± 0.08 0.35
        104 25.28 ± 0.06 0.24
        103 28.70 ± 0.09 0.31
        102 32.13 ± 0.19 0.58
        10 35.63 ± 0.45 1.27
    HTLV-3
        105 22.15 ± 0.06 0.27
        104 25.59 ± 0.09 0.37
        103 29.12 ± 0.06 0.21
        102 32.67 ± 0.26 0.80
        10 35.37 ± 0.46 1.29
    β-Globin
        105 23.17 ± 0.05 0.23
        104 26.51 ± 0.06 0.22
        103 29.97 ± 0.12 0.41
        102 33.46 ± 0.23 0.68
        10 36.85 ± 0.80 2.18
Interassay reproducibility
    HTLV-1
        105 22.72 ± 0.26 1.13
        104 26.94 ± 0.47 1.75
        103 30.52 ± 0.52 1.70
        102 34.08 ± 0.56 1.65
        10 37.70 ± 1.60 4.25
    HTLV-2
        105 23.70 ± 0.14 0.58
        104 27.30 ± 0.21 0.75
        103 30.83 ± 0.22 0.70
        102 34.28 ± 0.27 0.77
        10 38.38 ± 1.45 3.79
    HTLV-3
        105 24.33 ± 0.16 0.65
        104 28.11 ± 0.21 0.74
        103 31.66 ± 0.31 0.98
        102 35.23 ± 0.56 1.56
        10 39.52 ± 2.00 5.00
    β-Globin
        105 21.58 ± 0.18 0.82
        104 24.88 ± 0.18 0.74
        103 28.37 ± 0.19 0.68
        102 31.84 ± 0.25 0.77
        10 35.61 ± 0.69 1.95