Published in final edited form as: J Sep Sci. 2009 May;32(10):1632–1641. doi: 10.1002/jssc.200900074

Table 2.

Elution time of peak maximum (tp) and slopes of peak decay curves for racemic warfarin on the HSA column and control columns at 25°C and various flow ratesa

Flow rate
(ml/min)
1 µM Warfarin
on HSA column
10 µM Warfarin
on HSA column
10 µM Warfarin
on control column
tp (s) Slope (s−1) tp (s) Slope (s−1) tp (s) Slope (s−1)
0.1 231.3 −0.015 (± 0.001) 139.6 −0.020 (± 0.001) 82.4 −0.051 (± 0.001)
0.5 48.6 −0.061 (± 0.002) 29.2 −0.086 (± 0.005) 17.1 −0.182 (± 0.002)
1.0 25.7 −0.094 (± 0.006) 16.3 −0.147 (± 0.012) 10.0 −0.443 (± 0.012)
1.5 17.8 −0.150 (± 0.007) 11.9 −0.196 (± 0.002) 7.5 −0.551 (± 0.009)
2.0 14.2 −0.162 (± 0.007) 9.4 −0.229 (± 0.009) 6.2 −0.666 (± 0.007)
2.5 12.0 −0.190 (± 0.006) 8.2 −0.250 (± 0.006) 5.1 −0.855 (± 0.006)
3.0 9.9 −0.212 (± 0.014) 7.3 −0.344 (± 0.003) 4.8 −0.915 (± 0.013)
3.5 9.1 −0.236 (± 0.022) 6.7 −0.347 (± 0.007) 4.5 −0.981 (± 0.015)
4.0 8.0 −0.248 (± 0.019) 6.0 −0.344 (± 0.007) 4.2 −1.226 (± 0.013)
a

The values in parentheses represent a range of ± 1 S.D. All experiments in this table were performed using a 100 µl sample volume.