Table 4.
One-way ANOVA test results for each FFR model
| FFR model | Penetration (P-value) | Resistance (P-value) |
| N95 FFRs | ||
| N95-Aa | 0.0635 | 0.1233 |
| N95-Ba | 0.5761 | 0.0035b |
| N95-Ca | 0.8067 | 0.7572 |
| Surgical N95 FFRs | ||
| SN95-Da | 0.7688 | 0.0170b |
| SN95-Ec | 0.2189 | 0.2448 |
| SN95-Fa | 0.9409 | 0.0014b |
| P100 FFRs | ||
| P100-Ga | 0.2185 | 0.7446 |
| P100-Ha | 0.3046 | 0.4970 |
| P100-Ic | 0.0438a | 0.2580 |
For each FFR model with the exceptions of SN95-E and P100-I, one-way ANOVAs compare observed filter aerosol penetration or filter airflow resistance values for six test treatments [five different decontamination treatments and one as-received (control) group].
Values in bold font are P-value <0.05. Probability (P-value) of observing the given F-statistic or larger by chance.
The one-way ANOVAs compare observed filter aerosol penetration and filter airflow resistance values for five test treatments [four different decontamination treatments and one as-received (control) group]. Respirator component materials melted for these FFRs during microwave oven irradiation and subsequently samples were not evaluated for initial filter aerosol penetration and initial filter airflow resistance.