Table 4.

Stability of quality control samples (n = 5)

Nominal concentration
(ng/mL)
Measured mean concentration
(ng/mL) ± S.D
CV(%) Mean accuracy (%)
 Three freeze/thaw cycles (−20 °C)
20 18.38 ± 0.63 3.45 91.96
500 577 ± 27.03 4.69 115.50
2000 2202 ± 87.01 3.95 108.75
 Autosampler, 4 °C, 72 h
20 20.40 ± 1.26 6.45 102.12
500 493 ± 38.70 7.85 98.56
2000 1960 ± 175.78 8.97 98.14
 Ambient 24 h
20 17.4 ± 0.83 4.80 87.00
500 508.75 ± 11.98 2.36 101.93
2000 1840 ± 85.44 4.64 92.10
 Storage stability (−20 °C; 1 month)
20 18.66 ± 0.94 5.28 93.28
500 464.80 ± 20.58 4.43 92.90
2000 1874 ± 64.27 3.43 93.70