Skip to main content
. Author manuscript; available in PMC: 2010 Nov 15.
Published in final edited form as: Am J Cardiol. 2009 Sep 26;104(10):1343–1350. doi: 10.1016/j.amjcard.2009.07.003

Table 7.

Anterior Air Noise, Descending Aorta Noise, and Tube Current. SD denotes standard deviation, CI denotes confidence interval, HU denotes Hounsfield units.

Mean SD 95% CI P Value
Anterior Air Noise (HU) Sequential-if-appropriate Strategy vs. 16.8 3.7 16.3 – 17.3 0.0019
Helical-only Strategy 15.5 4.5 14.7 – 16.2

Sequential Protocol vs. 17.1 3.8 16.6 – 17.6 <0.0001
Helical Protocol 15.4 4.3 14.8 – 16.0

Helical Protocol in Sequential-if-appropriate Strategy vs. 15.1 3.2 14.0 – 16.2 0.62
Helical Protocol in Helical-only Strategy 15.5 4.5 14.7 – 16.2

Descending Aorta Noise (HU) Sequential-if-appropriate Strategy vs. 36.7 8.5 35.7 – 37.8 <0.0001
Helical-only Strategy 32.1 8.2 30.7 – 33.4

Sequential Protocol vs. 37.4 8.4 36.3 – 38.5 <0.0001
Helical-only Protocol 32.2 8.1 30.9 – 33.4

Helical Protocol in Sequential-if-appropriate Strategy vs. 32.5 7.9 29.8 – 35.1 0.79
Helical Protocol in Helical-only Strategy 32.1 8.2 30.7 – 33.4

Tube Current (mA) Sequential-if-appropriate Strategy vs. 445.3 122.3 430.4 – 460.2 0.0044
Helical-only Strategy 477.4 67.7 466.0 – 488.8

Sequential Protocol vs. 433.7 116.3 418.4 – 449.0 <0.0001
Helical Protocol 485.7 87.1 472.7 – 498.8

Helical Protocol in Sequential-if-appropriate Strategy vs. 517.8 135.2 472.0 – 563.5 0.013
Helical Protocol in Helical-only Strategy 477.4 67.7 466.0 – 488.8