Skip to main content
. 2009 Oct 28;16(12):1781–1788. doi: 10.1128/CVI.00248-09

TABLE 3.

Precision results of IgG anti-PT ELISAa

Parameter %CV for sample (geometric mean concn [EU/ml])
S1 (49) S2 (58) S3 (83) S4 (100) S5 (135) S6 (204) S7 (217)
Intra-assay precision
    Analyst 1 22 25 18 15 24 15 21
    Analyst 2 16 12 8 10 13 15 13
    Analyst 3 12 10 15 10 14 14 13
    Analyst 4 12 12 11 12 14 12 14
    Analyst 5 16 16 18 21 17 19 18
    Analyst 6 17 14 13 13 16 12 27
Interassay precision
    Site A 22 23 22 24 23 23 26
    Site B 21 29 23 21 25 21 26
    Site Bb 16 17 15 16 18 19 22
Reproducibility
    Site-to-site 23 28 23 23 27 25 29
    Site-to-sitec 20 22 20 21 23 23 27
a

Samples are ordered from low to high concentrations. Boldface (samples 2 to 5) indicates those samples with an IgG anti-PT concentration in the range analyzed under the prespecified acceptance criteria (50 to 200 EU/ml).

b

Interassay precision without analyst 1 from site B.

c

Reproducibility without analyst 1 from site B.