Skip to main content
. 2009 Oct 28;16(12):1781–1788. doi: 10.1128/CVI.00248-09

TABLE 4.

IgG anti-PT ELISA accuracy results obtained from analytical validation

Anti-PT IgG concn (EU/ml) for CBER3 (reference) Anti-PT IgG concn (EU/ml)
% Accuracy
Replicate 1 Replicate 2 Replicate 3 Geometric mean Expected value
50 52 50 41 47 50 95
100 85 92 81 86 100 86
150 120 124 113 119 150 79
200 142 139 151 144 200 72