. 2009 Oct 21;47(12):3920–3926. doi: 10.1128/JCM.01288-09

TABLE 4.

Real-time PCR limit of detection by the use of DNA extracted from spiked swabs

Extraction methoda LOD using DNA from indicated swabb
Polyester Polyurethane foam Rayon
MagNA Pure Compact 104 (50) 105 (500) 104 (50)
IT 1-2-3 105 (500) 105 (500) 106 (5,000)
MasterPure 104 (50) 104 (50) 105 (500)
QIAamp 105 (500) 106 (5,000) 105 (500)
UltraClean 105 (500) 105 (500) 104 (50)
a

Extraction methods were performed on samples recovered from triplicate swabs spiked with 10-fold dilutions of B. melitensis at a starting concentration of 107 CFU/ml.

b

The limit of detection (LOD) was determined to be the lowest concentration at which three out of three replicates produced a positive result (CFU/ml [CFU/reaction]).