Skip to main content
. 2009 Nov;3(6):1270–1281. doi: 10.1177/193229680900300606

Table 3.

Characteristics of Glucose Meters Useda

Test method Enzyme Specimen Sample volume (μl) Linearity (mg/dl) Analysis time (s)
Reference method YSI 2300 Amperometric Glucose oxidase Whole bloodb 25 0–900 90c
LifeScan
 One Touch II Hospital Photometric Glucose oxidase Whole blood 10 0–600 15–30
 SureStep Pro/Flexx Photometric Glucose oxidase Whole Blood 10 0–500 15–30
Abbott/MediSense
 Precision QID Amperometric Glucose oxidase Whole blood 3.5 20–600 20
 Precision PCx Amperometric Glucose oxidase Whole blood 3.5 20–600 20
Roche
 Accu-Chek Advantage Amperometric Glucose dehydrogenase Whole blood 4 10–600 26
 Accu-Chek Comfort Curve Amperometric Glucose dehydrogenase Whole blood 4 10–600 26
Bayer
 Elite XL Amperometric Glucose oxidase Whole blood 2 20–600 30
a

Reprinted with permission from Archives of Pathology & Laboratory Medicine. Copyright 2006. College of American Pathologists.67

b

Can also use plasma and serum.

c

Sample to sample