Skip to main content
. Author manuscript; available in PMC: 2010 Dec 1.
Published in final edited form as: Ophthalmology. 2009 Sep 10;116(12):2305–2314.e2. doi: 10.1016/j.ophtha.2009.05.025

Table 3.

Repeatability of Diagnostic Parameters

N PPG PG
Diagnostic
Parameters
ICC SD CV ICC SD CV ICC SD CV
RTVue FD-OCT
GCC-AVG (µm) 0.98 1.03 1.09 0.99 1.06 1.23 0.99 0.98 1.25
GCC-FLV (%) 0.91 0.38 0.96 0.59 0.95 1.01
GCC-GLV (%) 0.98 0.68 0.99 0.90 0.99 1.01
GCC-PCV 0.85 0.01 0.92 0.01 0.93 0.01
GCC-SID (µm) 0.94 1.12 0.95 1.24 0.97 1.61
MR-AVG (µm) 0.99 1.20 0.52 0.99 1.06 0.49 0.99 1.36 0.65
Stratus TD-OCT
NFL-AVG (µm) 0.96 1.71 1.72 0.99 1.52 1.75 0.98 2.22 2.86
MR-AVG (µm) 0.97 2.17 0.91 0.93 3.69 1.61 0.96 3.01 1.36

Abbreviations: N = normal; PPG = pre-perimetric glaucoma; PG = perimetric glaucoma; ICC = intraclass correlation; SD: Pooled standard deviation; CV = coefficient of variation. GCC = ganglion cell complex thickness; AVG = average; FLV (%) = focal loss volume expressed as a percentage of the average GCC volume in the normal group; GLV = global loss volume; PCV = pattern coefficient of variation; SID = superior - inferior difference; MR = macular retina thickness; NFL = nerve fiber layer thickness. FD-OCT: Fourier-domain optical coherence tomography; TD-OCT: time-domain optical coherence tomography