TABLE 2.
Troubleshooting table.
Step | Problem | Possible reason | Solution |
---|---|---|---|
18 | 50 μl of medium does not stay on coverslips | The coverslips do not recover to room temperature and are still wet | Leave the plates at room temperature longer or leave the lid half open for quick evaporation |
19 | The coverslips dry out | The drop of medium has drained off the coverslip | When moving and transferring the plate, make sure this is done gently. Ensure that the incubator door is opened and closed gently |
22 | No PDGFRα stained | The cells are overfixed or not fixed properly | Stain for the membrane protein before fixation as described in the notes of Table 1 |
23 | No 04+ cells | 04 cells died and detached; or inappropriate staining procedure | Adding 0.5% (vol/vol) fetal bovine serum into the culture medium may help the survival of 04+ cells. Incubate the cells with 04 antibody for 15 min before fixing the cells with 4% PFA (wt/vol) |
PDGFRα, platelet-derived growth factor receptor alpha; PFA, paraformaldehyde.