TABLE II.
Simulation results when random C-arm calibration and pose estimation errors are introduced. The results for each seed density are averaged over the source angle separations varying from 10° to 25° with a step size of 5° (total 800 simulations). The averaged results are rounded up to one digit after the decimal point which would be enough for clinically meaningful data analysis.
| Num of seeds |
Num of images |
Number of seeds | Mean ± STD error (mm) |
|
|---|---|---|---|---|
| candidate | correctly detected | |||
| 54 | 3 | 57.6 | 53.7 (99.4%) | 0.8 ± 0.5 |
| 4 | 54.3 | 54.0 (100.0%) | 0.8 ± 0.3 | |
| 60 | 3 | 65.4 | 59.7 (99.5%) | 0.8 ± 0.4 |
| 4 | 60.5 | 59.9 (99.8%) | 0.8 ± 0.2 | |
| 72 | 3 | 80.5 | 71.3 (99.0%) | 0.9 ± 0.6 |
| 4 | 72.9 | 71.9 (99.9%) | 0.8 ± 0.3 | |
| 84 | 3 | 96.4 | 82.7 (98.5%) | 0.9 ± 0.5 |
| 4 | 85.5 | 83.7 (99.6%) | 0.8 ± 0.3 | |
| 96 | 3 | 108.3 | 93.9 (97.8%) | 0.9 ± 0.6 |
| 4 | 100.8 | 95.3 (99.3%) | 0.8 ± 0.3 | |
| 112 | 3 | 128.8 | 108.3 (96.7%) | 0.9 ± 0.7 |
| 4 | 118.9 | 110.6 (98.8%) | 0.8 ± 0.4 | |