TABLE I.

Counts of MDA-231 Human Breast Cancer Cells in the Dielectric-Affinity Column Before and After Separation

Target Ratio Measured starting Ratio Tumor cells on electrodes before Separation Tumor cells on electrodes after Separation Efficiency of purging tumor cells
1:3 1:2.9 N/A N/A >95%
1:103 1:1.1×103 N/A N/A >95%
1:104 1:1.2×104 98 94 96%
1:104 1:1.8×104 56 53 95%
1:105 1:0.9×105 3 3 100%
1:3×105 1:2.8×105 1 1 100%

: Applied voltage = 2 V rms at 50 kHz; eluate flow rate = 5 μL/min; suspension conductivity =10 mS/m.

: The in-situ counting of tumor cells in the separation column was made for the tumor to normal cell ratio higher than 1:104. Below that, the counting involved too many cells and purging efficiencies were estimated by counting residual tumor cells that contaminated the “supposed” blood cell fraction.