Skip to main content
. 2009 Nov 17;36(12):5622–5632. doi: 10.1118/1.3259729

Figure 4.

Figure 4

Crossplane profiles measured for the 6 MV beam with the flattening filter out. Normalization is arbitrary. Depths are 1.5, 10, 20, and 30 cm. Lines are measured and points are simulated. Top: Difference between simulated and measured results. Solid circles represent a difference in percent, and open circles represent a difference in millimeters.