. 2009 Nov 30;9:48. doi: 10.1186/1471-2466-9-48

Table 2.

Results of statistical analysis on repeatability

EcoScreen
Volume pH LTB4 NOx 8-iso PGF PGE2

bias % 3.5 0.2 +) 48.9 2.1 +)
CV % 7.7 0.9 166.3 21.4

r2 0.479 0.889 n.c. 0.816
p 0.027 < 0.0001 n.c. 0.002
ICC 0.692 0.943 n.c. 0.903

ECoScreen2

Volume pH LTB4 NOx 8-iso PGF PGE2

bias % 0.5 0.7 9.0 +) 3.2 11.9
CV % 8.0 1.0 18.6 11.5 34.4

r2 0.618 0.915 0.656 0.626 0.704
p 0.021 0.0002 0.015 0.017 0.018
ICC 0.786 0.957 0.810 0.791 0.839

Repeated measurements of volume, pH and the mediators in EBC obtained with either the ECoScreen or ECoScreen2 device within 30 minutes. Values of the mean differences of sampling period 1 and 2 (bias), CVs and correlations. CV, coefficient of variation; ICC, intraclass correlation coefficient; n.c., not calculated. +)mediator concentration < LOQ.