Table 3.

Regression models for the post-mortem retrievals (adjusted R2=0.71, p<0.0001, RMS error=0.174 MPa) and laboratory-prepared specimens (adjusted R2=0.49, p=0.0022, RMS error=0.992 MPa) relating the dependent variable tensile strength of the cement-bone interface and independent variables including morphology parameters contract fraction (CF), intersection fraction (IF), and cement-bone gap thickness (GT).

Regression Model Term Estimate Standard error of mean P value
Post-Mortem Retrievals

   Constant (b1) −0.066 0.072 0.365
   CF (b2) 0.884 0.261 0.0015
   IF (b3) 0.223 0.082 0.0090
   GT (b4) (mm) 0.040 0.110 0.717
Laboratory-Prepared
   Constant (b1) 2.93 0.561 <0.0001
   CF (b2) 2.64 0.651 0.0008
   IF (b3) −1.5555 0.521 0.0083
   GT (b4) (mm) −4.52 10.32 0.667