. 2009 Dec;4(12):1988–1992. doi: 10.2215/CJN.03970609

Table 1.

Absolute values obtained for Dry and EDTA samples at baseline and after 1, 3, 6, and 12 mo of storage at −80°C on Liaison and Elecsys

Sample Liaison
Dry
EDTA
Baseline 1 mo 3 mo 6 mo 12 mo Baseline 1 mo 3 mo 6 mo 12 mo
1 226 194 165 160 130 240 194 172 191 199
2 807 695 648 743 764 831 685 613 653 652
3 766 661 602 677 691 690 578 516 519 568
4 331 248 238 267 287 334 256 217 229 245
5 133 103 97 109 105 116 100 82 93 97
6 728 553 534 605 642 723 587 509 546 543
7 473 387 364 391 430 468 370 349 324 380
8 448 375 358 403 435 438 340 272 279 351
9 291 246 218 251 264 299 249 206 208 253
10 440 381 320 373 409 451 397 313 352 416
11 339 231 261 292 277 337 254 243 249 269
12 198 168 163 178 174 195 165 163 166 174
13 461 423 400 431 458 483 427 399 421 467
14 182 157 152 157 177 176 162 149 145 160
15 53 46 46 44 39 53 48 43 40 43
16 61 51 49 54 58 57 51 44 47 54
Sample Elecsys
Dry
EDTA
Baseline 1 mo 3 mo 6 mo 12 mo Baseline 1 mo 3 mo 6 mo 12 mo
1 220 213 217 189 217 218 188 202 209 202
2 1244 1162 1200 1179 1190 1239 1185 1122 1099 1023
3 738 687 693 694 671 659 606 562 558 585
4 343 309 322 319 330 337 304 296 295 297
5 119 109 111 114 111 110 102 98 100 99
6 1212 1090 1163 1130 1120 1226 1113 1098 1061 1009
7 744 688 720 707 703 723 672 665 652 662
8 586 545 569 564 546 590 524 480 482 487
9 352 344 346 341 338 340 319 311 299 305
10 449 412 426 423 408 446 409 406 391 395
11 297 276 282 280 277 296 277 285 283 277
12 209 203 209 209 199 209 196 207 205 200
13 420 392 411 384 411 412 395 397 399 391
14 196 182 190 193 191 188 182 187 187 181
15 63 63 64 69 61 61 60 61 64 59
16 56 54 54 57 56 55 53 50 55 54