Published in final edited form as: J Memb Sci. 2010 Feb 15;348(1-2):131–149. doi: 10.1016/j.memsci.2009.10.050

Table 4.

Repetitive determination of hydraulic properties for the same CMA/12 probe (4•2) with a 4-mm 100-kDa MWCO polyethersulfone membrane used in the solute transport measurements in Fig. 10. The collection vial apparent elevation (Δhapp) was varied at fixed perfusate flow rates (Qin). Time lapsed refers to the approximate time since the first measurement was performed.

Lapsed time Qin Δhapp eff Hm κm ×10+13
μL/min cm
cmH2O·minμL
μLmin·cm2·cmH2O
cm2
-- 1 −26, −13, −1, +11, +24 17.6 0.48 2.5
1 week 1 −26, −13.5, −1, +11, +24 21.0 0.49 2.6
6 weeks 2 −26, −1, +12 4.4 1.16 6.2
7 weeks 1 −26, −9, −1 27.0 1.42 7.5

Mean±SD 17.5±9.6 0.89±0.47 4.7±2.5