System start up |
- Upon power up the red lasers turn on and move to home position |
Functional |
|
- Operator station (OS) computer turns on and operator logs in |
|
|
- System initializes, couch and laser positions are displayed on position control panel |
|
System shut down |
- Operator station computer turned off, key turned off on the machine and switched off. Red lasers turn off and the power distribution unit is turned off |
Functional |
Manual couch control |
- Couch moves up or down and in or out only when the enable button underneath is pressed |
Functional |
DICOM import |
- Import of DICOM images (kVCT) and structures to the system and display at the operator and planning station (PS) |
Images can be viewed at OS and PS |
Safety interlock |
- Mode switch: Imaging MVCT can not be formed when the mode switch is at treat |
Functional |
|
- Visual and audible indicators working |
|
|
- Door interlock working – checked during MVCT imaging procedure |
|
|
- Stop button on position control panel (PCP) working |
|
|
- Each emergency stop inside and outside the treatment room and on status control shuts down power to the gantry |
|
Laser alignment |
Verify home positions for movable red lasers where they overlap the fixed overhead green lasers within ±1 mm |
Verified |
|
Create a MVCT scan procedure for the cylindrical cheese phantom with the dowel at 5 mm position out. Red lasers move from home position to 2, 8 and 4 cm in the X, Y and −Z directions, respectively. |
|
|
Dose during MVCT scan is less than 4 cGy |
|
|
MVCT images are reconstructed and viewable from Reconstruction tab. |
|
|
Lateral and vertical shift for the reference “cross” on the MVCT image is ±1 mm and 5±1 mm, respectively. |
|
Treatment delivery test |
Treatment plans for “on axis” and “off axis” targets on a cheese phantom are delivered. The measured dose at each point is within ± 3% at ± 3 mm of the calculated point. |
Verified |
Interrupted treatment completion test |
Two films are exposed sequentially with uninterrupted and interrupted but completed procedures. Visual inspection of the films shows no discontinuity and point dose difference is within 3%. |
Verified |