Table 2.

Current Biomechanical Assessments for Dental Implant Osseointegration

Methodology Destructive Clinical use Property
Investigated
Parameters Reference
Tensional test Yes No Lateral resistance Maximal lateral load (Brånemark et al., 1998,
Kitsugi et al., 1996)
Push-out/Pull-out Yes No Interfacial shear Maximal force
Interfacial stiffness
(Berzins et al., 1997,
Brunski et al., 2000)
Removal Torque Yes No Interfacial shear Loosening torque
Torque load
(Johansson et al., 1998,
Meredith et al., 1997)
Cutting Resistance/Insertional
Torque
No Yes Interfacial shear Peak insertional torque
Torque load
(Friberg et al., 1995,
O’Sullivan et al., 2000)
Periotest No Yes Damping Periostest value (PTV) (Aparicio et al., 2006,
Schulte and Lukas, 1993)
Resonance Frequency Analysis No Yes Vibration/Damping Implant stability quotient
(ISQ)
(Friberg et al., 1999,
Meredith et al., 1997,
Turkyilmaz et al., 2009)