Skip to main content
. 2008 May 20;35(6):2414–2423. doi: 10.1118/1.2912177

Table 2.

Summary of exposure parameters resulting in the highest FOM value for each FFDM system, for each of the nine breast types tested. Parameters are in the form target∕filter∕kVp∕HVL, with kVp in kV and HVL in mm Al.

  Thickness (cm) 30∕70 50∕50 70∕30
  3 Mo∕Rh∕27∕0.43 Mo∕Rh∕27∕0.43 Mo∕Rh∕27∕0.43
GE 5 Mo∕Rh∕27∕0.43 Mo∕Rh∕27∕0.43 Mo∕Rh∕27∕0.43
  7 Mo∕Rh∕27∕0.43 Rh∕Rh∕31∕0.48 Rh∕Rh∕33∕0.53
         
  3 W∕Rh∕27∕0.49 W∕Rh∕27∕0.49 W∕Rh∕29∕0.51
Siemens 5 W∕Rh∕27∕0.49 W∕Rh∕27∕0.49 W∕Rh∕29∕0.51
  7 W∕Rh∕27∕0.49 W∕Rh∕29∕0.51 W∕Rh∕29∕0.51
         
  3 Mo∕Rh∕27∕0.39 Mo∕Rh∕27∕0.39 Mo∕Rh∕27∕0.39
Hologic 5 Mo∕Rh∕27∕0.39 Mo∕Rh∕27∕0.39 Mo∕Rh∕27∕0.39
  7 Mo∕Rh∕27∕0.39 Mo∕Rh∕28∕0.40 Mo∕Rh∕28∕0.40
         
  3 W∕Al∕27∕0.42 W∕Al∕27∕0.42 W∕Al∕27∕0.42
Fischer 5 W∕Al∕29∕0.45 W∕Al∕30∕0.46 W∕Al∕30∕0.46
  7 W∕Al∕35∕0.54 W∕Al∕35∕0.54 W∕Al∕35∕0.54
         
  3 Mo∕Mo∕24∕0.30 Mo∕Mo∕24∕0.30 Mo∕Mo∕26∕0.32
Fuji 5 Mo∕Mo∕24∕0.30 Mo∕Mo∕24∕0.30 Mo∕Mo∕26∕0.32
  7 Mo∕Rh∕28∕0.43 Mo∕Rh∕28∕0.43 Mo∕Rh∕28∕0.43