Table 2.
Summary of exposure parameters resulting in the highest FOM value for each FFDM system, for each of the nine breast types tested. Parameters are in the form target∕filter∕kVp∕HVL, with kVp in kV and HVL in mm Al.
| Thickness (cm) | 30∕70 | 50∕50 | 70∕30 | |
|---|---|---|---|---|
| 3 | Mo∕Rh∕27∕0.43 | Mo∕Rh∕27∕0.43 | Mo∕Rh∕27∕0.43 | |
| GE | 5 | Mo∕Rh∕27∕0.43 | Mo∕Rh∕27∕0.43 | Mo∕Rh∕27∕0.43 |
| 7 | Mo∕Rh∕27∕0.43 | Rh∕Rh∕31∕0.48 | Rh∕Rh∕33∕0.53 | |
| 3 | W∕Rh∕27∕0.49 | W∕Rh∕27∕0.49 | W∕Rh∕29∕0.51 | |
| Siemens | 5 | W∕Rh∕27∕0.49 | W∕Rh∕27∕0.49 | W∕Rh∕29∕0.51 |
| 7 | W∕Rh∕27∕0.49 | W∕Rh∕29∕0.51 | W∕Rh∕29∕0.51 | |
| 3 | Mo∕Rh∕27∕0.39 | Mo∕Rh∕27∕0.39 | Mo∕Rh∕27∕0.39 | |
| Hologic | 5 | Mo∕Rh∕27∕0.39 | Mo∕Rh∕27∕0.39 | Mo∕Rh∕27∕0.39 |
| 7 | Mo∕Rh∕27∕0.39 | Mo∕Rh∕28∕0.40 | Mo∕Rh∕28∕0.40 | |
| 3 | W∕Al∕27∕0.42 | W∕Al∕27∕0.42 | W∕Al∕27∕0.42 | |
| Fischer | 5 | W∕Al∕29∕0.45 | W∕Al∕30∕0.46 | W∕Al∕30∕0.46 |
| 7 | W∕Al∕35∕0.54 | W∕Al∕35∕0.54 | W∕Al∕35∕0.54 | |
| 3 | Mo∕Mo∕24∕0.30 | Mo∕Mo∕24∕0.30 | Mo∕Mo∕26∕0.32 | |
| Fuji | 5 | Mo∕Mo∕24∕0.30 | Mo∕Mo∕24∕0.30 | Mo∕Mo∕26∕0.32 |
| 7 | Mo∕Rh∕28∕0.43 | Mo∕Rh∕28∕0.43 | Mo∕Rh∕28∕0.43 |