. 2008 Apr 8;35(5):1682–1697. doi: 10.1118/1.2889778

Table 2.

Mean and standard deviation of reconstructed μa and μs in mm−1 values for the background and a spherical target with no scattering contrast using the LM and GLS techniques. The diameter of the spherical inclusion was 15 mm. Data with 1% noise were used.

  Target position Background Target
Methods   μa μs μa μs
Actual 0.01 1.0 0.02 1.0
LM (0,0,0) 0.0101±0.0003 1.0025±0.0211 0.0109±0.0001 1.0500±0.0071
(30,0,0) 0.0101±0.0004 1.0009±0.0205 0.0119±0.0004 1.0934±0.0269
GLS (0,0,0) 0.0101±0.0001 1.0016±0.0269 0.0126±0.0002 1.0924±0.0253
(30,0,0) 0.0100±0.0004 1.0029±0.0189 0.0136±0.0003 1.1002±0.0294