Table 1.
cm/s | Scanner Site |
Aorta | MPA (HF phase-enc) |
MPA (LR phase-enc) |
---|---|---|---|---|
Scanner type 1 | 1 | -2.2 cm/s | -3.8 cm/s | -5.6 cm/s |
2 | -2.8 | -3.1 | -5.5 | |
3 | -0.7 | -3.2 | -5.3 | |
4 | -1.3 | -3.5 | -4.7 | |
Scanner type 2 | 5 | -2.2 | -3.4 | -5.4 |
6 | -1.1 | -3.9 | -3.1 | |
7 | 1.6 | -4.0 | -3.6 | |
8 | -2.7 | -4.9 | -4.9 | |
Scanner type 3 | 9 | 1.6 | 1.2 | 1.8 |
10 | 1.2 | -0.7 | 0.4 | |
11 | 0.9 | 1.6 | 2.1 | |
12 | 1.1 | -1.7 | 2.0 | |
Mean of all 12 absolute values | 1.6 cm/s | 2.9 cm/s | 3.7 cm/s | |
Range of all 12 absolute values | 0.7 to 2.8 cm/s | 0.7 to 4.9 cm/s | 0.4 to 5.6 cm/s |
The twelve rows are from the four sites using each scanner type. The scanner types are not identified.