Table 3.

Reproducibility of macular thickness measurements by scan type.*

Stratus Macular Thickness Cirrus Cube 512×128 RTVue MM5 RTVue MM6 3D OCT-1000Radial 3D OCT-1000 3D Macular
ICC* SD ICC SD ICC SD ICC SD ICC SD ICC SD
Fovea 92.02 3.32 97.25 1.91 88.54 3.44 90.75 1.76 97.88 1.10 96.59 0.71
Parafovea
 Temporal 94.93 2.15 92.61 2.44 82.64 2.60 61.09 0.00 96.41 0.00 82.06 0.00
 Superior 94.63 1.67 95.30 2.10 92.07 2.31 92.72 2.19 94.96 1.28 95.87 1.21
 Nasal 84.53 2.80 94.85 2.13 93.73 1.69 92.6 0.00 93.7 1.13 94.61 1.73
 Inferior 93.99 1.67 94.34 1.86 76.43 3.09 92.14 1.14 95.26 1.18 92.28 0.00
Perifovea
 Temporal 89.38 1.98 93.28 1.56 82.44 0.00 91.94 2.01 93.11 0.00 31.47 2.43
 Superior 92.17 2.42 94.41 2.31 80.12 0.00 96.78 0.00 94.73 1.71 95.71 1.43
 Nasal 91.16 1.19 96.93 1.91 92.2 2.76 85.41 0.00 95.25 0.00 97.46 1.45
 Inferior 84.8 0.00 95.55 1.41 88.82 2.47 95.36 1.22 94.33 1.15 64.38 0.00
*

ICC = intraclass correlation coefficient expressed as % (p<0.01 for each ICC)

SD = intervisit standard deviation expressed in μm.