Table 5.
Scan Type | Signal Strength* | Any Artifact | Inaccurate CFT† | IRM‡ | ORM** | Out of Register | Off Center | Cut Edge |
---|---|---|---|---|---|---|---|---|
Stratus Macular Thickness | 8.83±1.22 | 17.46% | 4.76% | 12.70% | 0.00% | 1.59% | 6.35% | 6.35% |
Cirrus Cube 512×128 | 9.93±0.42 | 43.75% | 0.00% | 15.00% | 36.25% | 8.75% | 0.00% | 0.00% |
RTVue MM5 | 61.76±9.10 | 32.50% | 6.25% | 22.50% | 1.25% | 7.50% | 11.25% | 0.00% |
RTVue MM6 | 62.72±9.74 | 38.75% | 8.75% | 35.00% | 7.50% | 0.00% | 6.25% | 0.00% |
3D OCT-1000 Radial | 60.54±9.29 | 28.75% | 15.00% | 26.25% | 1.25% | 5.00% | 5.00% | 0.00% |
3D OCT-1000 3D Macular | 50.64±9.95 | 53.16% | 17.72% | 34.18% | 12.66% | 18.99% | 7.59% | 0.00% |
Signal strengths are determined by automated software on each instrument.
CFT = central subfield foveal thickness
IRM = inner retinal misidentification
ORM = outer retinal misidentification.