Table 5.

Artifacts and signal strength of all scans acquired in first visit.

Scan Type Signal Strength* Any Artifact Inaccurate CFT IRM ORM** Out of Register Off Center Cut Edge
Stratus Macular Thickness 8.83±1.22 17.46% 4.76% 12.70% 0.00% 1.59% 6.35% 6.35%
Cirrus Cube 512×128 9.93±0.42 43.75% 0.00% 15.00% 36.25% 8.75% 0.00% 0.00%
RTVue MM5 61.76±9.10 32.50% 6.25% 22.50% 1.25% 7.50% 11.25% 0.00%
RTVue MM6 62.72±9.74 38.75% 8.75% 35.00% 7.50% 0.00% 6.25% 0.00%
3D OCT-1000 Radial 60.54±9.29 28.75% 15.00% 26.25% 1.25% 5.00% 5.00% 0.00%
3D OCT-1000 3D Macular 50.64±9.95 53.16% 17.72% 34.18% 12.66% 18.99% 7.59% 0.00%
*

Signal strengths are determined by automated software on each instrument.

CFT = central subfield foveal thickness

IRM = inner retinal misidentification

**

ORM = outer retinal misidentification.