Table 2.

Comparison of performance of different types of detectors. The following detector specifications were used: reach-through SPAD: SPCM AQR-14 (Perkin Elmer Optoelectronics, Fremont, CA, USA); epitaxial SPAD: PDM 50CTC (Micro Photon Devices, Bolzano, IT); GaAsP PMT: H7422-40 (Hamamatsu Photonics, Bridgewater, NJ, USA); CCD: CoolSnap HQ (Photometrics, Tucson, AZ, USA), ICCD: I-Pentamax (Princeton Instruments, Trenton, NJ, USA); EMCCD: Cascade 512B (Photometrics)

H33D Single-Photon Counting Detectors Imagers
Gen I Gen III Reach-through
SPAD
Epitaxial
SPAD
GaAsP PMT CCD ICCD EMCCD
Technology Vacuum Vacuum Solid State Solid State Vacuum Solid State Vacuum + Solid State Solid State
Photon Counting + + + + +
Sensitive material S20 GaAs Si Si GaAsP Si GaAs Si
QE (@ 600 nm) 4% 35-40% 70% 30% 40% 60% 45% 90%
Time Resolution 100 ps 250 ps 300 ps 50 ps 400 ps 100 ms 60 ms 34 ms
Global Rate* 500 kHz 20 MHz N/A N/A N/A 217 GHz 315 MHz 12 GHz
Local Rate* 10 kHz 100
kHz
10 MHz 10 MHz 10 MHz 150 kHz 1.2 kHz 46 kHz
Nb of Pixels π 1282 π 2562 1 1 1 1392 ×1040 512 × 512 512 × 512
Pixel Size 100 μm 35 μm 170 μm 50 μm 5 mm 6.5 μm 15 μm 16 μm
Max Frame Rate 1 kHz 1 kHz ¼ frame/min ¼ frame/min ¼ frame/min 10 Hz 15 Hz 29 Hz
*

For imagers, the local count rate is replaced by the pixel well capacity divided by the maximum gain (electrons per photon) divided by the maximum frame rate. Using the minimum gain would allow obtaining higher count rates, but would result in worse single-molecule detection (lower SNR) because of the larger influence of the readout noise.

Better timing resolution (down to a few ms) can be achieved by limiting the acquisition to a small number of pixels, or by binning pixels. Notice also that ICCD’s can in some case be gated to obtain ~ 200 ps nanotime-resolution (see text).

Frame rates for point detectors are calculated for a raster-scanned image of 512 × 512 pixels, and assuming a dwell-time of 1 ms/pixel.