OS-ADS method. Deviations among the full FOV, small FOV, and extended small FOV data in each of the nine myocardial segments calculated for the worst-case scenario of simultaneous 99mTc/201Tl SPECT

Segment Deviation (%)
Small FOV Extended by 0.5 FWHM Extended by 1 FWHM Extended by 1.5 FWHM
Total 7.8 3.0 0.75 −0.43
Apex 6.8 2.5 0.68 0.10
Anterior mid 2.0 −0.40 −1.2 −1.3
Anterior basal 10 3.8 2.6 2.9
Lateral mid 6.2 3.0 1.4 0.023
Lateral basal 7.8 4.1 2.2 −0.65
Inferior mid 7.9 2.2 −0.68 −1.2
Inferior basala 49 15 1.8 −1.4
Septal mid 4.0 0.72 −1.7 −2.1
Septal basala 23 5.5 1.4 −1.0

aSegments containing perfusion defect.