. 2009 Dec 4;80(6):646–652. doi: 10.3109/17453670903487008

Table 4.

The prevalence of metal allergies and the type of material inserted during hip operation in patients who were patch-tested prior to the operation (n = 292) and in patients who were patch-tested after the operation (n = 64)

Cases patch-tested
before operation (n = 292) % (n) after operation (n = 64) % (n) OR (95% CI) a p-value a
Metal allergy b 16 (47) 6.3 (4) 2.87 (1.00–8.30) 0.4
Nickel allergy 11 (33) 4.7 (3) 2.59 (0.77–8.73) 0.1
Cobalt allergy 6.5 (19) 0 (0) 0.03
Chromium allergy 3.1 (9) 3.1 (2) 0.98 (0.21–4.67) 1.0
Cause of revision
 Aseptic loosening 7.5 (22) 9.4 (6) 0.79 (0.31–2.03) 0.2
 Osteolysis 0 1.5 (1)
 Deep infection 1.0 (3) 1.5 (1) 0.65 (0.07–6.39)
 Fracture or luxation 1.3 (4) 0
 Other 0.3 (1) 0
Femoral head material
 Metal 81 (236) 84 (54) 0.78 (0.37–1.63) 0.6
 Ceramics 13 (38) 13 (8) 1.05 (0.46–2.37)
 Not exchanged (at revision) 0 0
 Other or missing 6.2 (18) 3.1 (2) 2.04 (0.46-9.01)
Acetabular liner material
 Polyethylene 35 (103) 13 (8) 3.81 (1.75–8.31) 0.001
 Metal 1.7 (5) 0
 Ceramics 1.0 (3) 0
 Other or missing 62 (162) 88 (56) 0.18 (0.08–0.39)

a Comparison of cases and controls (Chi-squared test). Fisher's exact test was used when the expected frequency in one of the cells was lower than 5.

b Metal allergy was defined as a positive patch-test reaction to nickel, cobalt or chromium.

OR: odds ratio.

CI: confidence interval.