. 2010 Jan;4(1):145–150. doi: 10.1177/193229681000400118

Table 3.

Accuracy of Blood Glucose Meters as Compared to YSI Reference Values in ISO 15197 Table Format

YSI reference value <75 mg/dl YSI reference value ≥75 mg/dl Overall acceptable results (within ±15 mg/dl and ±20%)
Within ±5 mg/dl Within ±10 mg/dl Within ±15 mg/dl Within ±5% Within ±10% Within ±15% Within ±20%
System A (2/4) 50% (3/4) 75% (4/4) 100% (56/99) 56.6% (83/99) 83.8% (94/99) 94.9% (97/99) 98.0 % (101/103) 98.1%
System B (0/4) 0% (2/4) 50% (4/4) 100% (54/97) 55.7% (77/97) 79.4% (89/97) 91.7% (93/97) 95.9% (97/101) 96.0%
System C (2/4) 50% (2/4) 50% (4/4) 100% (34/98) 34.7% (68/98) 69.4% (90/98) 91.8% (95/98) 96.9% (99/102) 97.1%
System D (1/3) 33.3% (2/3) 66.7% (3/3) 100% (38/85) 44.7% (57/85) 67.1% (74/85) 87.1% (80/85) 94.1% (83/88) 94.3%