Skip to main content
. 2010 Jan 27;37(2):842–847. doi: 10.1118/1.3298015

Table 1.

Values of normalized CTDIw at the widest available beam width, for eight commercial CT scanners. N∕A: Not available.

Vendor Scanner model CTDIw (mGy∕100 mAs)
80 kV 100 kV 120 kV 140 kV
GE HiSpeed CT∕i 1.6 3.6 5.2 7.6
LightSpeed Plus 2.9 5.6 9.0 12.9
LightSpeed 16 2.7 5.8 9.1 13.7
VCT 3.4 6.2 9.5 13.3
Siemens Somaton Plus 4 2.3 N∕A 7.9 11.4
Sensation 4 2.5 N∕A 7.7 10.9
Sensation 16 2.4 4.8 7.6 10.9
Sensation 64 1.8 4.1 7.0 11.4
Average ±σ 2.5±0.6 5.0±1.0 7.9±1.4 11.5±1.9