Table 3.

Optimized radiographic technique for FFDM using W–Rh target∕filter determined from FOM results of the current study using an investigational a-Se FFDM system (Siemens Mammomat NovationDR, Siemens Medical System). The results reported are for masses and calcifications (calcification results are shown in parentheses).

Optimized radiographic technique for FFDM using W–Rh spectrum
Breast thickness (cm) Breast phantom composition
Adipose (kVp) Mixed (50–50) (kVp) Glandular (kVp)
2 26 (27) 28 (28) 29 (29)
4 27 (28) 29 (29) 29 (29)
6 29 (29) 29 (29) 29 (29)
8 30 (30) 30 (30) 30 (29)