Table 3.
Cylinder activity quantification
Cylinder volume (ml) | b | Recovery coefficient (from spheres) | Activity without RC (MBq) | Activity with RC (MBq) |
---|---|---|---|---|
50 | 0 | 0.78 | 0.56 (−79%) | 0.72 (−39%) |
100 | 0 | 0.86 | 0.67 (−50%) | 0.77 (−29%) |
200 | 0 | 1.00 | 0.85 (−18%) | 0.85 (−18%) |
50 | 1/5 | 0.87 | 0.70 (−44%) | 0.80 (−25%) |
100 | 1/5 | 0.90 | 0.76 (−32%) | 0.84 (−19%) |
200 | 1/5 | 1.00 | 0.91 (−9%) | 0.91 (−9%) |
50 | 1/3 | 0.93 | 0.78 (−28%) | 0.84 (−19%) |
100 | 1/3 | 0.93 | 0.82 (−22%) | 0.88 (−13%) |
200 | 1/3 | 1.00 | 0.96 (−4%) | 0.96 (−4%) |
The values in parentheses are the percentage difference between the estimated activity and the true value of 1 MBq. VOIs were defined using physical boundaries