Skip to main content
. Author manuscript; available in PMC: 2010 Mar 5.
Published in final edited form as: Nat Protoc. 2008;3(8):1287–1298. doi: 10.1038/nprot.2008.119

Table 5.

Limits of detection, DNA requirements and parameters for preparation of calibration curve samples

Lesion LOD (fmol) Minimum DNA (μg) Unlabeled standard (nM) Internal standard (nM)
dX 10 50 62.5-10001 250
dU 50 50 62.5-1000 250
dI 5 10-20 6.25-100 25
8-oxo-dG 5 50 6.25-100 25
1,N2-εdG 10 100 0.625-10 2.5
εdC 0.5 50 0.625-10 2.5
εdA 1 10-20 0.625-10 2.5
1

Prepare 5 concentrations differing by 2-fold (e.g., 62.5, 125, 250, 500, 1000)