Skip to main content
. Author manuscript; available in PMC: 2010 Mar 5.
Published in final edited form as: Nat Protoc. 2008;3(8):1287–1298. doi: 10.1038/nprot.2008.119

Table 8.

Troubleshooting table

Problem Possible reason(s) Solution(s)
Low DNA yields → Incomplete lysis of tissue or cells → Optimize cell or tissue disruption conditions (e.g., increase time or intensity of mechanical homogenization)
→ DNA yield is sometimes tissue dependent, some tissues yield higher DNA amount than others → Increase sample size

RNA contamination → Inadequate RNA degradation during DNA isolation → Increase RNase during extraction and prolong the incubation time

Inconsistent retention times during HPLC pre-purification or LC/MS-MS → Column not equilibrated prior to purification → Increase column equilibration time
→ Inconsistent column pressures due to inadequate degassing of mobile phases → Use a degassing apparatus on the HPLC or Sonicate buffers for 20 min
→ Problems with the stationary phase in the column (e.g., use of high aqueous isocratic conditions during LC/MS/MS analysis leads to column collapse) → Regenerate column by running 100% acetonitrile overnight at 0.100 ml/min; replace column, if necessary

dI levels are higher than expected → Deaminase contamination → Include or increase coformycin during all stages of DNA isolation and processing
→ Contamination of dI with dA during HPLC pre-purification or LC/MS-MS; m/z values differ by 1, so dA isotopomer signals (M+1) contribute to dI signal → Decrease the % organic solvent to resolve the dA and dI peaks

dX and dO levels are higher than expected → Contamination of dX or dO with dG during HPLC pre-purification or LC/MS-MS; m/z values differ by 1, so dG isotopomer signals (M+1) contribute to dX/dO signal → Decrease the % organic solvent to resolve the dG and dX or dO peaks

8-Oxo-dG levels are higher than expected → Artifactual formation during sample work-up or in vacuo drying → Include desferrioxamine and butylated hydroxytoluene during cell processing, DNA extraction and hydrolysis steps; process samples on ice; minimize sample evaporation time
→ As 8-oxo-dG can be formed from 2′-deoxyguanosine (dG) in the ion source of the mass spectrometer it is important that dG contamination separates well from 8-oxo-dG prior to entering the MS → Decrease the % organic solvent to resolve dG from 8-oxo-dG

εdC levels are higher than expected → Contamination of εdC with dA due to similar HPLC retention times and identical m/z values → Optimize the separation of εdC and dA in the pre-purification step. Also, optimize the HPLC step of the LC/MS-MS analysis to such that εdC elutes 2-3 min before dA.