Table 1.

Result Summary (all measurements except exposure time are expressed in millimeters)

Canine Number Treatment Time Atrial Septal Axial Motion ASD
Hemorrhage
Ultrasound (Center)* Pathology Pathology
RA side LA side RA side LA side
1 6 min 5.2 1.7 3.4 2.9 1.2 1.9
2 10 min 17.1 No ASD was created
3 8 min 8.4 3.3 5.1 2.2 2.3 2.1
4 6 min 10.3 5.0 5.7 2.4 4.2 2.2
5 10 min 10.0 4.2 4.7 2.6 1.6 2.3
6 6 min 8.7 3.0 6.6 3.0 1.6 2.0
7 6 min 9.7 3.0 4.5 2.2 2.3 1.7
8 6 min 6.5 3.2 6.2 2.5 3.2 2.4
9 16 min 6.1 4.0 No pathology result§
10 12 min 8.1 3.6 4.1 3.3 4.0 2.6
mean 8.4 min 8.1 3.4 5.0 2.6 2.6 2.2
std 3.4 min 3.3 0.9 1.1 0.4 1.1 0.3
*

”ASD Ultrasound (Center)” column reports the width of the color jet in the center of the ASD measured on Ultrasound Doppler images.

”ASD Pathology RA side” and “LA side” columns show the widths of the acellular zone in the ASD for the right and left atrial surfaces, respectively, measured based on pathology analysis using H&E cross-section slides.

”Hemorrhage Pathology RA side” and “LA side” report the width of the eosinophilic and hemorrhagic rim surrounding the acellular ASD zone for the right and left atrial surfaces, respectively. This width is measured from the outermost edge of the acellular zone to the outermost edge of eosinophilic myocyte and hemorrhage zone on either side of the ASD. For each ASD, this measurement was performed for both sides of the ASD and using multiple slides across the same ASD, and the maximal value of the measurements is reported.

§

Pathological evaluation on the heart of dog #9 could not be obtained because it was filled with heart worms. However, ultrasound Doppler confirmed the ASD creation.