Skip to main content
. Author manuscript; available in PMC: 2010 Aug 1.
Published in final edited form as: Nat Protoc. 2010 Feb 4;5(2):357–370. doi: 10.1038/nprot.2009.246

Table 2. Troubleshooting table.

Step Problem Possible reason Solution
3, 14, 15, 17 Fluorochrome
‘spillover’ (i.e.
non-specific
fluorescence
detected in a
neighboring
channel, e.g.
false-positive
events in the PE
channel
attributable to
FITC
fluorescence)
  1. Improper selection of fluorochrome, or inadequate understanding of antigen and epitope distribution.

  2. Improper instrument setup and fluorochrome compensation.

  3. Use of tandem dyes (e.g. PE-Cy5).

Check fluorescence spectrum viewer
(http://bdbiosciences.com/spectra or
http://www.invitrogen.com/site/us/en/home/support/Research-Tools/Fluorescence-SpectraViewer.html). Examine single-
color staining profiles and perform
‘fluorescence minus one’ or FMO36
control stainings to identify the culprit
fluorochrome. Knowing the
distribution of antigens and epitopes
targeted by antibodies, and having a
proper understanding of
fluorochromes, their selection and
instrument setup and operation are
crucial for polychromatic flow
cytometry10,34-39. Refer also to
instrument user’s guide provided by
the flow cytometer manufacturer.
3B, 17 Dull staining of
manually
conjugated
Pacific Blue
antibody (e.g.
anti-CD123) using
the Invitrogen
Zenon labeling kit
 i. Suboptimal antibody: conjugate ratio. Optimize amount of Component A
used, and titrate antibody levels.
Component B (blocking
immunoglobulin) must also be
adjusted stoichiometrically (refer to
manufacturer’s instructions).
Alternatively, a Pacific Blue equivalent
such as eFluor 450 can be used.
7, 11, 17 Lack of
consistency in
phenotype
readout.
  1. Donor variation or instability of phenotype.

  2. Lack of a consistent time interval between cell fixation and flow cytometric acquisition.

  3. Use of non-standardized, non-fresh reagents.

Allow samples to stabilize in fixative
for at least 1 h after immunostaining.
Analyze at a fixed time interval (e.g. 16
h after cell fixation). Avoid analysing
later than 24 h as fluorescence
intensity or tandem dyes might begin
to degrade. Use freshly prepared
standardized reagents (e.g. BD FACS
lysing solution, CellFIX).
7, 17 Odd or variable
FSC vs SSC
bivariate profiles
or large amounts
of debris,
including dead
cells.
  1. Prolonged incubation (>15 min) with FACS Lysing Solution can alter cellular biophysical properties or surface epitopes.

Stop incubation promptly at 10 min,
and invert tubes less vigorously.
9, 11, 17 Low cell yield.
  1. Discarding supernatant without care.

  2. Rarity of cell population.

Pipet off supernatant carefully if using
microcentrifuge tubes, or pour away
supernatant gently when using FACS
tubes. Consider scaling up blood
volumes and staining reagents.
14, 17 Progressive
shifting of
acquisition
events on FSC vs
SSC bivariate plot
over days and
sometimes during
a data collection
session. This can
  1. Obstruction of internal tubings that supply or drain flow cell.

  2. Fluidics or pressure problems in tubing circuit.

Ferromagnetic valves are often used
to regulate pressures and buffer or
waste flow within certain flow
cytometers. Use of magnetic bead-
based cell sorting and subsequent
analysis on such flow cytometers can
cause valves to malfunction, leading
to increased internal pressure within
the flow cell. Discontinue the use of