Table 2. Troubleshooting table.
Step | Problem | Possible reason | Solution |
---|---|---|---|
3, 14, 15, 17 | Fluorochrome ‘spillover’ (i.e. non-specific fluorescence detected in a neighboring channel, e.g. false-positive events in the PE channel attributable to FITC fluorescence) |
|
Check fluorescence spectrum viewer (http://bdbiosciences.com/spectra or http://www.invitrogen.com/site/us/en/home/support/Research-Tools/Fluorescence-SpectraViewer.html). Examine single- color staining profiles and perform ‘fluorescence minus one’ or FMO36 control stainings to identify the culprit fluorochrome. Knowing the distribution of antigens and epitopes targeted by antibodies, and having a proper understanding of fluorochromes, their selection and instrument setup and operation are crucial for polychromatic flow cytometry10,34-39. Refer also to instrument user’s guide provided by the flow cytometer manufacturer. |
3B, 17 | Dull staining of manually conjugated Pacific Blue antibody (e.g. anti-CD123) using the Invitrogen Zenon labeling kit |
i. Suboptimal antibody: conjugate ratio. | Optimize amount of Component A used, and titrate antibody levels. Component B (blocking immunoglobulin) must also be adjusted stoichiometrically (refer to manufacturer’s instructions). Alternatively, a Pacific Blue equivalent such as eFluor 450 can be used. |
7, 11, 17 | Lack of consistency in phenotype readout. |
|
Allow samples to stabilize in fixative for at least 1 h after immunostaining. Analyze at a fixed time interval (e.g. 16 h after cell fixation). Avoid analysing later than 24 h as fluorescence intensity or tandem dyes might begin to degrade. Use freshly prepared standardized reagents (e.g. BD FACS lysing solution, CellFIX). |
7, 17 | Odd or variable FSC vs SSC bivariate profiles or large amounts of debris, including dead cells. |
|
Stop incubation promptly at 10 min, and invert tubes less vigorously. |
9, 11, 17 | Low cell yield. |
|
Pipet off supernatant carefully if using microcentrifuge tubes, or pour away supernatant gently when using FACS tubes. Consider scaling up blood volumes and staining reagents. |
14, 17 | Progressive shifting of acquisition events on FSC vs SSC bivariate plot over days and sometimes during a data collection session. This can |
|
Ferromagnetic valves are often used to regulate pressures and buffer or waste flow within certain flow cytometers. Use of magnetic bead- based cell sorting and subsequent analysis on such flow cytometers can cause valves to malfunction, leading to increased internal pressure within the flow cell. Discontinue the use of |