Skip to main content
. Author manuscript; available in PMC: 2011 Feb 1.
Published in final edited form as: IEEE Trans Med Imaging. 2010 Feb;29(2):474–481. doi: 10.1109/TMI.2009.2036843

Table III.

Relative Errors and Correlation Coefficients of Reconstructed Electric Properties without and With HZ Component Using Tem and Birdcage Coils

REs / CCs 28cm TEM without HZ 28cm BC without HZ 28cm BC with HZ 12cm BC without HZ
CSF σ 27.1% 0.971 30.1% 0.899 28.2% 0.943 39.4% 0.8525
εr 24.9% 0.936 26.3% 0.902 24.3% 0.920 38.1% 0.8908
WM σ 21.4% 0.940 28.6% 0.898 25.1% 0.908 38.9% 0.8430
εr 12.7% 0.982 15.3% 0.963 12.6% 0.987 28.6% 0.9258
GM σ 28.6% 0.913 32.2% 0.888 27.7% 0.919 40.6% 0.8513
εr 18.0% 0.969 18.8% 0.958 16.5% 0.974 39.1% 0.8585