TABLE 2.

Interassay concordance in the analysis of blood donor samples (n = 1455) from Chaco Province, Argentina (2006–2007)*

EIA BioM EIA BiosCh EIA BioZ EIA Wrec PA TIA IHA IHA LC
EIA BioM
EIA BiosCh 98.1
EIA BioZ 98.4 97.7
EIA Wrec 97.0 97.3 96.6
PA 95.2 96.3 95.8 95.1
TIA 95.1 94.4 95.1 94.4 93.1
IHA 92.1 93.6 92.7 92.6 94.6 90.4
IHA LC 88.0 89.6 89.0 88.3 91.6 87.1 93.3
*

For each pair of assays compared, concordance was defined as (Number positive by both assays + Number negative by both assays)/ Number tested (= 1455).

EIA BioM = Chagatek; EIA BiosCh = Chagas BiosChile; EIA BioZ = BioZima Chagas; EIA Wrec = Chagatest rec. 3.0; PA = Chagas Serodia; TIA = trans-sialidase inhibition assay; IHA = Chagas HAI; IHA LC = HAI LC.