Table 1.

Gamma Camera Performance (Tc-99m)

Parameter Value
Spatial resolution 2.2 mm FWHM (all source directions)
Energy resolution 10.8% (at 140 keV)
Sensitivity 149 cps/MBq (± 10% energy window)
Count rate linearity r = 0.996, slope = 149 cps/MBq, straight line fit
Spatial uniformity < ± 3% variation across UFOV