Exposure Time (seconds) | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
60 | 90 | 120 | 150 | 180 | ||||||||||||
W* | L† | V‡ | W | L | V | W | L | V | W | L | V | W | L | V | ||
Power (W) | ||||||||||||||||
10 | Mean | 11.6 | 11.9 | 848 | 14.3 | 14.5 | 1556 | 17.1 | 16.6 | 2532 | 17.8 | 18.5 | 3113 | 20.4 | 22.3 | 4821 |
SD£ | 0.9 | 0.2 | 145 | 0.3 | 2.1 | 275 | 1.2 | 0.1 | 356 | 1.2 | 3.8 | 863 | 0.1 | 1.8 | 350 | |
12.5 | Mean | 13.3 | 13.3 | 1249 | 16.2 | 17.0 | 2382 | 20.3 | 21.0 | 4529 | 20.5 | 21.0 | 4676 | 21.2 | 22.2 | 5197 |
SD | 0.8 | 1.0 | 244 | 1.3 | 2.2 | 606 | 1.1 | 1.4 | 776 | 2.1 | 1.4 | 1269 | 1.8 | 1.4 | 735 | |
15 | Mean | 15.5 | 18.3 | 2341 | 18.4 | 21.0 | 3718 | 21.1 | 23.3 | 5462 | 22.1 | 24.2 | 6230 | 24.6 | 25.8 | 8214 |
SD | 1.5 | 1.6 | 650 | 2.1 | 1.6 | 816 | 1.9 | 2.2 | 1204 | 1.7 | 2.2 | 1464 | 1.6 | 1.6 | 1521 |
W = width (mm)
L = length (mm)
V = volume (mm3)
SD = standard deviation
Total of 57 ablations, range of 2 to 6 repeat measurements for each setting