Skip to main content
. Author manuscript; available in PMC: 2011 Feb 1.
Published in final edited form as: J Orthop Res. 2010 Feb;28(2):218–224. doi: 10.1002/jor.20971

Table 1.

Failure forces for OrthoCoupler and control specimens at 60 days post-surgery (p = 0.0003; Student t test for paired data).

Failure forces (N) Paired data
Goat Control OrthoCoupler Control OrthoCoupler

1 61 247 61 247
2 115 213 115 213
3 95 268 95 268
4 48 186 48 186
5 185 437 185 437
6 --- 387 -- --
7 219 437 219 437
Mean 120.5 310.7 120.5 298
SD 68.3 107.0 68.3 111.3