Skip to main content
. 2010 Mar 23;5:12. doi: 10.1186/1749-8546-5-12

Table 2.

Intra- and interday variability for the assay of the three Panax notoginseng saponins in beagle dog plasma (n = 6)

Spiked conc.
(ng/mL)
Intraday Interday


Measured conc. (ng/mL) RSD (%)a Accuracy (%)b Measured conc. (ng/mL) RSD (%) Accuracy (%)
Notoginsenoside R1
2.8 2.8 1.27 100.0 2.68 1.49 95.71
14 13.9 0.76 99.29 12.89 9.62 92.1
140 148 8.22 105.7 131.67 10.11 105.7
Ginsenoside Rg1
3.8 3.6 5.13 94.7 3.96 11.96 104.2
19 17.6 3.65 92.6 18.3 8.95 96.3
190 197.3 6.59 103.8 193.7 7.66 101.7
Ginsenoside Rb1
18 18.1 3.07 100.6 17.6 5.36 96.11
180 165.3 2.73 91.83 179.7 5.05 99.50
900 912.67 3.90 101.3 871.3 2.38 96.87

Note: aRelative standard deviation; bAccuracy (%) = [1--(nominal concentration -- mean of measured concentration)/nominal concentration] × 100