Skip to main content
. 2010 Feb 10;48(4):1408–1412. doi: 10.1128/JCM.02463-09

TABLE 1.

Mean cycle threshold (CT) values for replicates of prepared BD GeneOhm StaphSR assay positive- and negative-control reagents frozen at −70°C for 3 weeks, 9 weeks, and 18 weeksa

Reconstituted reagent Mean CT ± SD following −70°C storage for:
3 wks 9 wks 18 wks
Positive controlb 35.15 ± 0.29 35.20 ± 0.24 35.40 ± 0.22
Negative controlc 33.87 ± 0.43 34.10 ± 0.40 34.59 ± 0.16
a

CT values for negative-control material are indicative of internal-control amplification.

b

The mean CT value ± standard deviation for freshly prepared positive-control reagent (22 determinations) was 34.67 ± 0.48.

c

The mean CT value ± standard deviation for freshly prepared negative-control reagent (22 determinations) was 34.63 ± 0.32.